Feasibility of a new
hollow fiber silicone membrane oxygenator for long-term
ECMO application
Shinji Kawahito, Tomohiro Maeda,
Tadashi Motomura, Tamaki Takano,
Kenji Nonaka, Joerg Linneweber, Seiji Ichikawa, Hiroshi
Ishitoya,
Kazuhiro Hanazaki, Julie Glueck, Koshiro Sato*, and Yukihiko
Nose
|
Michael E. DeBakey Department of Surgery, Baylor
College of Medicine, Houston, Texas, USA;and *Fuji Systems,
Inc., Tokyo, Japan
Abstract: Currently in United States, there are no clinically-applicable
hollow fiber extracorporeal membrane oxygenation (ECMO) oxygenators
available. Therefore, our laboratory is in the process of
developing a silicone hollow fiber membrane oxygenator for
long-term ECMO usage. This oxygenator incorporates an ultrathin
silicone hollow fiber. At this time, a specially-modified
blood flow distributor (one chamber distributor) is centered
in the module to prevent blood stagnation. An ex vivo long-term
durability test for ECMO was performed using a healthy miniature
calf for 2 weeks. Venous blood was drained from the left jugular
vein of a calf, passed through the oxygenator and infused
into the left carotid artery using a Gyro C1E3 centrifugal
blood pump. A successful 2-week ex vivo experiment was performed.
The O2 and CO2 gas transfer rates were maintained at the same
value of 40 ml/min at a blood flow rate of 1L/min flow and
V/Q=3 (V=gas flow rate;Q=blood flow rate). The plasma free
hemoglobin was maintained around 5 mg/dl. After the experiment,
no blood clot formation was observed in the module and no
abnormal necropsy findings were found. These data suggest
that the performance of this newly-improved oxygenator was
stable, reliable, and acceptable for long-term ECMO. J. Med.
Invest. 49:156-162, 2002
Keywords:hollow fiber, silicone membrane, oxygenator, extracorporeal
membrane oxygenation, long-term ex vivo study
INTRODUCTION
An effective microporous hollow fiber oxygenator has been
developed for cardiopulmonary bypass procedures;however, since
1970, no improvements have been made for an effective extracorporeal
membrane oxygenation (ECMO) system (1). The Kolobow spiral
coil membrane oxygenator (Medtronic Inc., Anaheim, CA, U.S.A.)
was developed in 1972 (2) and is the only practical available
ECMO system in the United States. This oxygenator is an old-fashioned
device that is inefficient and difficult to handle.
To solve this problem, our laboratory is developing a new
oxygenator using ultrathin silicone rubber hollow fibers (3-7).
These devices have demonstrated better biocompatibility and
higher gas transfer performances over and above the already
existing ECMO oxygenator (6, 7). For further high performance
and antithrombogenicity, an oxygenator with an improved blood
flow distributor was fabricated and evaluated with in vitro
(8) and short term ex vivo (9) testing. The purpose of this
study was to evaluate its clinical long-term feasibility for
ECMO using an animal model.
METHODS
Animal preparation
The animal involved in this study received humane care in
compliance with the "Guiding Principles in the Case and
Use of Animals" approved by the Council of the American
Physiological Society (revised 1980) and the "Guide for
the Care and Use of Laboratory Animals" published by
the National Institutes of Health (NIH Publication No. 85-23,
revised 1985). A healthy miniature female calf (Dexter strain)
weighing 90 kg and 11 months old was the subject of this study.
A calf of this age was essential for this study because an
animal with normal physiological adult circulation was needed,
without abnormal anatomy due to the fetal circulation (i.e.
atrial septal defect). Complete blood cell count (CBC), blood
biochemistry, and plasma free hemoglobin were examined before
surgery.
Surgical procedure
Anesthesia was induced using 4% halothane with 50% nitrous
oxide through a special mask. When the animal was anesthetized,
endotracheal intubation was performed and anesthesia was maintained
by 1-2% halothane with 100% oxygen. A 15 cm longitudinal incision
was made along the jugular vein on the left side of the neck,
and the left carotid artery and jugular vein were dissected.
The arterial cannula was inserted through a small arteriotomy
and threaded proximally into the artery;the venous cannula
was inserted through the jugular vein in the proximal direction
in the same manner. After placing the arterial and venous
cannulae, both were externalized and connected to the extracorporeal
circuit, and the pump circuit was activated. The Gyro C1E3®
(Kyocera Corporation, Kyoto, Japan) was used for the centrifugal
pump. Mean arterial pressures were maintained at more than
80 mmHg. Activated clotting time (ACT) was maintained at approximately
400 sec throughout the experiment using heparin. When the
calf completely recovered from the anesthesia, the ECMO study
was started. At the end of experiment, euthanasia and necropsy
were performed.
Membrane Oxygenator
A new silicone hollow fiber membrane oxygenator (Preproduction
model PPM-04, Fuji Systems Inc., Tokyo, Japan) was fabricated
for long-term ECMO application based on the results of a previous
model (PPM-03) (8, 9). The major changes in the new improved
model (PPM-04) are an increase in the fiber length (from 100
to 150 mm) and the surface area (from 0.8 to 1.0m2) to increase
the gas transfer rate, a decrease in the packing density (from
45 to 40%) to decrease the pressure drop, and a specially-designed
blood flow distributor (from 4 chambers to 1 chamber) incorporated
into the center of the module to prevent blood stagnation.
Consequently, the new ECMO oxygenator module was 220 mm long
and contained in a silicone coated acrylic housing. The priming
volume of this module was 200 ml. This oxygenator was of the
extracapillary flow type, in which blood flows outside the
hollow fibers (Fig. 1).
Measurements
Gas transfer rate:At a blood flow rate of 1 L/min and V/Q=3
(V=gas flow rate;Q=blood flow rate), the O2 and CO2 gas transfer
rates were evaluated for 2 weeks (ECMO condition). Blood gas
samples, taken from the inlet and outlet sampling ports, were
analyzed every day using a System 1306 pH/blood gas analyzer
(Instrumentation Laboratories, Lexington, MA, U.S.A.). Three
samples were measured for each sampling time. The O2 content
and O2 transfer rate and the CO2 content and transfer rates
were calculated by the following standard formulas:
O2 content (Vol%)=(Hb×1.34×%O2 saturation)/100+PO2×0.003
O2 transfer rate (ml/min)=(CaO2 - CvO2)×blood flow
rate
Total CO2 (mmol/L)=HCO3 -+0.03×PCO2
CO2 transfer rate (ml/min)=22.4×(tCO2v - tCO2a)×blood
flow rate
where Hb is hemoglobin (g/dl), PO2 is oxygen partial pressure
(mmHg), CaO2 is arterial oxygen content (Vol%), CvO2 is venous
oxygen content (Vol%), the blood flow rate represents pump
flow rate (L/min), HCO3 - is plasma bicarbonate ion concentration
(mmol/L), PCO2 is CO2 partial pressure (mmHg), tCO2v is venous
total CO2 (mmol/L), and tCO2a is arterial total CO2 (mmol/L).
Hemolysis test:Plasma free hemoglobin was measured every day.
The method of measuring plasma free hemoglobin has been described
by Mizuguchi et al. (10).
Pressure drop measurement:Pressure differences between the
inflow and outflow ports were monitored to assess the pressure
drop in the oxygenator throughout the experiment using a pressure
monitor (Living Systems Instrumentation, Burlington, VT, U.S.A.).
CBC and biochemistry study:CBC, lactate dehydrogenase (LDH),
aspartate aminotransferase (AST), alanine aminotransferase
(ALT), blood urea nitrogen (BUN), creatinine, total billirubin
(T. Bil), total protein (TP), and electrolytes (Na, K, Cl)
were measured every two days postoperatively.
RESULTS
A2-week ex vivo experiment was successfully performed without
exchanging the oxygenator. The stable and reliable performance
of this new oxygenator was demonstrated for the entire experiment.
Gas Transfer Performance:Fig. 2 shows the results of gas exchange
performance tests. The O2 and CO2 gas transfer rates at a
blood flow rate of 1 L/min and V/Q=3 were maintained at 41.72±4.13
(mean±SD) ml/min and 40.97±14.49 ml/min,
respectively, for 2 weeks.
Hemolysis test:As shown in Fig. 3, the plasma free hemoglobin
was maintained at 5.50±2.20 mg/dl for 2 weeks.
Pressure drop study:The pressure drop in the blood chamber
was kept from 20 to 40 mmHg. A high pressure drop was sometimes
observed, however, this was not continuously observed (for
only 2 or 3 hours) (Fig. 4). The reason of this high pressure
drop was unknown.
CBC and biochemistry study:Table 1 shows the CBC and blood
biochemistry data throughout the experiment. All the mean
values were within normal limits except for anemia due to
bleeding towards the end of the experiment. Just after the
operation, abnormal levels of WBC, AST and TP were observed,
however, they were normalized gradually. Also, the platelet
count was maintained above200,000/µl.
Visual inspection:After the experiment, the visual inspection
of the oxygenator revealed no blood clot or thrombus formation
(Fig. 5).
Necropsy findings:There were no thromboembolic findings in
major organs including liver, kidney, spleen, intestine, etc.
in the macroscopic examination. However, a huge hematoma in
the chest cavity due to the high ACT level was observed.
DISCUSSION
The microporous membrane oxygenator that was introduced in
the early 1960s has been successfully used for cardiopulmonary
bypass during cardiac surgery. However, plasma leakage through
the pores into the gas phase is one of the limiting factors
for long-term use of these microporous membrane oxygenators
(11-13). The plasma leakage is thought to be a result of the
loss of the hydrophobic characteristics of the micropores
by the adsorption of protein, and that is promoted by an increase
in the outlet blood pressure (14). Several techniques have
been proposed to decrease the plasma leakage, including the
use of a heated humiditied gas (11, 12, 15), use of high gas
flow rate (15), and development of a more uniform and smaller
size porous membrane (16, 17). These techniques require additional
complex and expensive devices. In contrast, the nonporous
true membrane oxygenator can completely prevent plasma leakage
without any other hardware.
ECMO is now a standard treatment for respiratory failure refractory
to conventional pulmonary support techniques worldwide, and
many patients are treated with ECMO every year (18). The Kolobow
spiral coil membrane oxygenator (2), the only ECMO oxygenator
approved in the U.S.A., employs a solid silicone membrane.
However, because of the mechanically weak characteristics
of silicone, manufacturing of hollow fibers from silicone
elastomers is difficult and expensive. This author's group
developed a novel silicone material with sufficient mechanical
strength, and a fine silicone hollow fiber with a diameter
of 300 µm and wall thickness of 50 µm
(3). Utilizing this hollow fiber, an oxygenator with an improved
blood flow distributor was fabricated. After satisfactory
in vitro (8) and short-term ex vivo (9) performances were
achieved, a 2-week ex vivo experiment was successfully performed.
Other groups are also developing high-performance ECMO oxygenators,
for example, Takewa et al. (19) reported long-term ECMO experiments
using a heparin bonded oxygenator. On the other hand, our
oxygenator was made of silicone rubber with excellent biocompatibility.
This is the first all silicone rubber hollow fiber ECMO oxygenator
in the world.
One of the limitations of conventional silicone hollow fiber
compared with microporous membrane oxygenators is poor gas
permeability. We solved the problem to some extent by using
a novel fine ultrathin silicone fiber. In this ex vivo experiment,
the O2 and CO2 gas transfer rates were maintained at the same
value of 40 ml/min at a blood flow rate of 1 L/min flow and
V/Q=3 for 2 weeks. Stable and reliable gas transfer performance
under the ECMO condition is advantageous for long-term usage.
However, O2 and CO2 transfer rates are still limited comparing
the results of in vitro experiments (8). Further improvements
(further increase of the fiber length and the surface area,
etc.) are necessary. Additionally, the plasma free hemoglobin
levels and pressure drops were within acceptable levels, and
no abnormalities during the blood examination or necropsy
were observed. Excellent biocompatibility of our silicone
membrane oxygenator was proven in this ex vivo experiment.
The remaining problem is the bleeding complication due to
the anti-coagulant therapy. Because the preliminary study
cases developed oxygenator occlusion after several days, it
was decided that the ACT level be increased from 200-250 sec
to around 400 sec. However, in this study, a huge hematoma
in the chest cavity due to the high ACT level was observed.
During the next ECMO experiment, this group is proposing the
ACT level be in the range of 250-300 sec. These results indicate
that this oxygenator can be used for prolonged ECMO, although
further improvement in thrombo-resistant properties should
be achieved.
In conclusion, a 2-week ex vivo experiment with the silicone
membrane hollow fiber oxygenator was successfully completed.
Stable and reliable performance of this oxygenator was proven.
These data demonstrate the feasibility for long-term application
of this newly improved silicone hollow fiber oxygenator. However,
this manuscript is essentially the results of a single experiment,
which confirms the feasibility of our newly-developed oxygenator.
It is not a sufficient number of experiments to confirm the
advantages of our oxygenator. This oxygenator is still in
the development stage, and further experiments and improvements
will be necessary.
REFERENCES
1.Nosé Y, Motomura T, Kawahito S:Oxygenator-Artificial
Lung-Past, present, and future. ICAOT/ICMT Press, Houston,
2001
2.Kolobow T, Spragg RG, Pierce JE, Zapol WM:Extended term
(to 16 days) partial extracorporeal blood gas exchange with
the spiral membrane lung in unanesthetized lambs. ASAIO Trans
17:350-354, 1971
3.Funakubo A, Higami T, Sakuma I, Fukui Y, Kawamura T, Sato
K, Sueoka A, Nosé Y:Development of a membrane oxygenator
for ECMO using a novel fine silicone hollow fiber. ASAIO J
42:M837-M840, 1996
4.Niimi Y, Yamane S, Yamaji K, Tayama E, Sueoka A, Nosé
Y:Protain adsorption and platelet adhesion on the surface
of an oxygenator membrane. ASAIO J 43:M706-M710, 1997
5.Yamane S, Ohashi Y, Sueoka A, Sato K, Kuwana J, Nosé
Y:Development of a silicone hollow fiber membrane oxygenator
for ECMO application. ASAIO J 44:M384-M387, 1998
6.Maeda T, Iwasaki A, Kawahito S, Nakata K, Nonaka K, Linneweber
J, Schulte-Eistrup S, Takano T, Yoshikawa M, Sato K, Kuwana
J, Murabayashi S, Nosé Y:Preclinical evaluation
of a hollow fiber silicone membrane oxygenator for extracorporeal
membrane oxygenator application. ASAIO J 46:426-430, 2000
7.Kawahito S, Maeda T, Takano T, Nonaka K, Linneweber J, Mikami
M, Motomura T, Ichikawa S, Glueck J, Sato K, Kuwana J, Nosé
Y:Gas transfer performance of a hollow fiber silicone membrane
oxygenator:Ex vivo study. Artif Organs 25:498-502, 2001
8.Kawahito S, Maeda T, Motomura T, Takano T, Nonaka K, Linneweber
J, Mikami M, Ichikawa S, Kawamura M, Glueck J, Sato K, Nosé
Y:Development of a new hollow fiber silicone membrane oxygenator:In
vitro study. Artif Organs 25:494-498, 2001
9.Kawahito S, Maeda T, Motomura T, Takano T, Nonaka K, Linneweber
J, Ichikawa S, Kawamura M, Ishitoya H, Glueck J, Sato K, Nosé
Y:Preclinical evaluation of a new hollow fiber silicone membrane
oxygenator for pediatric cardiopulmonary bypass:Ex-vivo study.
Ann Thorac Cardiovasc Surg 8:7-11, 2002
10.Mizuguchi K, Damm GA, Aber GS, Bozeman RJ, Bacak JW, Svejkovsky
PA, Orime Y, Ohara Y, Naito K, Takai K, Makinouchi K, Takatani
S, Nosé Y:Does hematocrit affect in vitro hemolysis
test results? Preliminary study with Baylor/NASA prototype
axial flow pump. Artif Organs 18:650-656, 1994
11.Palder SB, Shaheen KW, Whittelessy GC, Nowlen TT, Kundu
SK, Klein MD:Prolonged extracorporeal membrane oxtgenation
in sheep with a hollow fiber oxygenator and a centrifugal
pump. Trans ASAIO 34:820-822, 1988
12.Mottaghy K, Oedekoven H, Starmans H, Muller B, Kashefi
A, Hoffmann B, Bohn S:Technical aspects of plasma leakage
prevention in microporous capillary membrane oxygenators.
ASAIO Trans 35:640-643, 1989
13.Montoya JP, Shanley CJ, Merz SI, Bartlett RH:Plasma leakage
through microporous membranes:Role of phospholipids. ASAIO
J 38:M399-M405, 1992
14.Tamari Y, Tortolani AJ, Lee-Sensiba KJ:Bloodless testing
for microporous membrane oxygenator failure:A preliminary
study. Int J Artif Organs 14:154-160, 1991
15.Kawak AM, Gaylor JDS, Simpson K, Kirkland M:An in vitro
technique to assess oxygenator potential for respiratory failure
therapies. Int J Artif Organs 14;234-238, 1991
16.Muramoto T, Tatebe K, Nogawa A, Hagiwara K, Seita Y, Fukasawa
H, Takahashi K, Satoh M, Inaba A, Takahashi A:Development
of a new microporous hollow fiber membrane for oxygenators.
Jpn J Artif Organs (in Japanese) 19;472-475, 1990
17.Hagiwara K, Innami K, Yokoyama K, Kitoh H, Nogawa A, Muramoto
T, Tatebe K, Seita Y, Fukasawa H:An approach to the microporous
hollow fiber for the ECMO oxygenator-micropore characteristics
of the gas exchange performance, plasma leakage, and hydrophilization
of the inner surface of fibers. Jpn J Artif Organs (in Japanese)
21;720-726, 1992
18.Ichiba S, Bartlett RH:Current status of extracorporeal
membrane oxygenation for severe respiratory failure. Artif
Organs 20;120-123, 1996
19.Takewa Y, Tatsumi E, Taenaka Y, Nakatani T, Masuzawa T,
Nishimura T, Nakamura M, Endo S, Takano H, Tanaka S:Long-term
evaluation of a heparin bonded artificial lung-Continuous
use up to 36 days. Jpn J Artif Organs (in Japanese) 27;379-383,
1998
Abbreviations:LCAT;lecithin:cholesterol acyltrasferase, the
CE/TC ratio;the cholesterol ester/total cholesterol ratio,
the C/P ratio;the free cholesterol/total phospholipids ratio,
HDL;high density lipoprotein, ESR;electron spin resonance,
I(5.10):a stearic acid analogue of spin probe
Received for publication June 27, 2002;accepted July 31, 2002.
Address correspondence and reprint requests to Shinji Kawahito,
M.D., Department of Anesthesiology, The University of Tokushima
School of Medicine, Kuramoto-cho, Tokushima 770-8503, Japan
and Fax:+81-88-633-7182.
|
| |